• Beta
Data Quality (Identifying Anomalies)
  • 04 Oct 2024
  • 2 Minutes to read
  • Contributors
  • Dark
    Light
  • PDF

Data Quality (Identifying Anomalies)

  • Dark
    Light
  • PDF

Article summary

The CML measurements to be used for the Corrosion Rates (CR) calculation are shown on the CML Measurement History tab of the Circuit. IMS will identify "bad" readings (anomalies). For example, IMS will always identify when the current thickness reading is greater than the previous reading, since normally an Equipment's wall thickness does not get thicker. The next sections explain how IMS flags unusual readings.

Growth Warning “W”

A thickness reading that is thicker than the previous reading, but not more than 10% thicker than the previous reading will be flagged with a "W" (Warning). This reading will still be used in the calculations. The user should always review growth warnings!

The "W" will be removed if the current thickness reading is marked as “Verified” (a Quality Indicator) and the thickness reading will still be used in the IMS calculations. (Quality Indicator = Ignored, Verify, or Good.)

A Growth Warning occurs when both:

A growth warning.

Growth Error “E”

A reading that is 10% thicker than the previous or base reading, or 15% thicker than any other “Good” reading will not be used in the IMS calculations. IMS will mark this thicker reading with an "E" (Error). You should always review/verify “erred” readings!

The "E" will be removed if the thickness reading is marked as “Verified”. IMS can now use this reading in the calculations since it is no longer considered a “growth” error reading. Usually when verifying an “erred” reading, you should also “ignore” the ones that are really causing the problem.

A Growth Error occurs when:

A growth error.

Time Error “T”

Short time spans between thickness readings usually create unreal CRs. When corrosion calculations result in short time intervals, the user must review and understand what caused the short inspection interval. When thickness readings are taken at an interval less than 6 months, the previous reading will never be used in the IMS calculations. IMS will mark the previous reading with a "T" (Time Error).

Note: The "T" will not be removed when the current thickness reading is marked as Verified.

Duplicate Warning “D”

When the same value is inputted twice it will be marked with a “D”.

A duplicate warning

Historical Warning “H”

A reading will be marked as historical when a new Baseline reading is entered.

A historical warning due to a new Baseline reading.

Reviewing/Verifying Error Readings

When an Error has occurred, you will need to verify the Error. Ideally you should then first verify the reading in the field. After verifying the reading in the field, you can fix the calculation in IMS.

To do this in IMS:

  • Open the Circuit Details Page (see IMS Details Page).

  • Go to the CML | CML Measurement History section.

  • Click the CML Measurement History tab (for more information on the tab see PEI Circuits - Details Page).

  • Search for the CML with the error (E tag).

Verify the error:

  • Right click on the Quality indicator (“I” \ “G” \ “V”, etc.).

  • Select “Verified” or “Ignore” from the dropdown.

  • Click the Save button to save the CML Measurement History.

Right click on the Quality indicator (“I” \ “G” \ “V”, etc.) and select the desired indicator from the dropdown.

Take Note

After verifying an Error, you will need to re-calculate and Approve the calc summary (see Calculating and Approving the Calc Summary).


Was this helpful? Click to add feedback comments

Changing your password will log you out immediately. Use the new password to log back in.
First name must have atleast 2 characters. Numbers and special characters are not allowed.
Last name must have atleast 1 characters. Numbers and special characters are not allowed.
Enter a valid email
Enter a valid password
Your profile has been successfully updated.
ESC

Eddy AI, facilitating knowledge discovery through conversational intelligence